840 Dissolved Oxygen Transmitter

Loop Powered Dissolved Oxygen Monitoring

The 840 Transmitter has been designed to provide accurate, highly reliable measurements in difficult highly fouling applications. The specifically designed sensor utilises a rugged, 100 micron membrane, in combination with a unique electrolyte to provide excellent long term stability, without the need for sensor replenishment or replacement.

The result is amazing - the 840 is so stable that it can measure accurately for years with practically no attention. Service intervals of five years have been reported! Maintenance is as simple as wiping the membrane clean and checking the calibration from time to time. If it should ever be necessary to replace the membrane, this can be done quickly, easily and at very little cost.

When combined with the EasyCal calibration is carried out in 2 minutes with the probe in water. As the sensor does not need to acclimatise in air, errors do not occur by hurried calibrations. As with all aspects of the 840 Transmitter the calibration routine is extremely simple and involves no tables or complicated software features.

To make installation as easy as possible and to help achieve the best possible results we recommend using the Pioneer Probe holder. This places the sensor approximately 300 mm below the surface away from the surface effects of rain, sunshine and fouling. The flexible joint in the mounting shaft allows the sensor to move, providing an anti-fouling and anti-ragging action.

Applications

  • Activated Sludge Plant
  • River Monitoring
  • Aquaculture

Features

  • Long Sensor Life
  • Low Maintenance
  • Easy to use

Technical Specification

Physical

Dimensions (hxwxd)Monitor: 121 x 119 x 60 mm
Sensor: 58 mm Diameter, 56 mm long
WeightMonitor: 1.0 kg
Sensor: 0.4 kg (inc 5 metres of cable)
Protection ClassElectronics IP65, Sensor IP68
Enclosure MaterialMonitor: ABS
Sensor: Polyoxymethylene
Cable EntriesIntegral Cable Gland
Cable Size3 core, 5 mm OD
Cable Length10 metres standard, 100 metres maximum

Electrical

Loop Power15 - 35VDC
Max Load50 ohm at 15 VDC, 1050 ohm at 35 VDC

Environmental Data

Operating Temperature Monitor -20 to 60ºC, Sensor 0-50ºC
Storage Temperature -20 to 60ºC
Location Indoor/Outdoor

Measurement Principle

PrincipleGalvanic Cell, Self Polarising
Temperature CompensationSelf Compensating

Measurement Characteristics

Range0-5 to 0-40 mg/l
0-50 to 0-400 %SAT
Response Time90% of end value in less than 120 seconds
Accuracy+/- 2% of measured value
Repeatability+/- 0.5% of measured value
FlowrateTypcially less than 10 cm/seconds

User Interface

DisplayLCD Display providing 3 1/2 Digits
Setupvia Dip Switches
Units of Measurementmg/l, ppm, %SAT