IR 'C' Range

Suspended Solids Sensors for Environmental Research

The IR 'C' version of Partech's trusted range of infrared based Suspended Solids and Turbidity sensors is aimed at applications in the environmental monitoring and research fields. It has been specifically developed to work with a wide range of data logging devices that are in common used in universities and research organisations. The sensor produces an output signal that can be calibrated either within the logger or via a spreadsheet to provide a linear response to the actual solids present in the water being monitored.

The sensors have provided data for a multitude of academic papers in applications from Glacial Streams in the Alps, through Tropical Rainforest Rivers in Borneo to Melt Water in Greenland and sediment transport in the Humber Estuary. Many theses and doctoral reports have been produced used the IR 'C' sensors.

Being exceptionally easy to deploy and having a very low power consumption there are no hidden costs to the purchase of these sensors. Partech are happy to put together a monitoring package including a logger, power source and weatherproof enclosure. These packages will allow data to be retrieved without visiting the site, keeping project costs down and allowing measurements over wide areas.

Applications

  • Glacial Meltwater
  • Silt Transportation
  • River Quality

Features

  • Self Contained sensor
  • Easy Logger Interface
  • No Training Required

Technical Specification

Physical

Dimensions (hxwxd)IR100: 95 x 160 x 25 mm
Dimensions (hxwxd)IR40: 95 x 100 x 25 mm
Dimensions (hxwxd)IR15/8: 95 x 75 x 25 mm
Weight0.25 to 0.35 kg
Protection ClassIP68
Enclosure MaterialBlack Epoxy Resin
Cable Entryvia Integral Gland
Cable Type4 mm OD, Polyurethane Coated
Cable Length10 metres standard ,100 metres maximum

Environmental Data

Operating Temperature 0-40ºC
Storage Temperature 0-50ºC
Location Outdoor

User Interface

Output0-5 mA
Shallow 'S' shaped curve will need linearisation against actual site samples prior to use

Power Supply

Voltage 12VDC

Measurement Characteristics

Response Time0.5 seconds
Accuracy+/- 1 cm of interface
ResolutionStandard cable marking every 1.0 metres
Measurement PrincipleLight Absorption
Wavelength960 nm

Sensor Ranges

IR100 SensorRange: 0-200 mg/l
Optical Path: 200 mm
IR40 SensorRange: 0-1,500 mg/l
Optical Path: 40 mm
IR15 SensorRange: 0-10,000 mg/l
Optical Path: 15 mm
IR8 SensorRange: 0-30,000 mg/l
Optical Path: 8 mm