Turbi-Tech 2000LA Sensors
Self Cleaning Sensor for Suspended Solids Monitoring

The Turbi-Tech 2000LA Sensor has been designed for use in aeration systems typically monitoring Mixed Liquor Suspended Solids (MLSS) also known as Activated Sludge where
solids are typically in the range of 1,500 to 3,500 mg/l. The sensor can also measure Returned Activated Sludge (RAS), Surplus Activated Sludge (SAS) and
Suspended Solids or Turbidity in any higher range application.
The large optical surface and sample volume combine to ensure that the sensor is providing information that is reliable and representative of the solids present
in the process. Deposits of fats and grease on the sensing area do not prevent the sensor from measuring unlike sensors with small optical surfaces. In addition
the sensor incorporates a self cleaning mechanism that ensures that the optical surfaces are kept clean at all times, the cleaning system is designed to avoid
problems with ragging and does not smear the optical system. The cleaning system ensures that manual intervention on a routine basis is not required, the sensor
should simply be checked as part of general site housekeeping

The Turbi-Tech 2000LA Sensor uses Infrared Light at 860 nm and operates using the Light Attenuation Principle. The cleaning mechanism uses a polyurethane wiper seal, backed by 2 Nitrile 'H' Rings that finish the cleaning process. The Nitrile seals can be exchanged for Viton if the process media dictates. The cleaning process is automatically initiated by the 7200 Monitor at a user determined frequency. The cleaning process takes only 90 seconds, which means that the sensor is available for 99.5% of the time with a 5 hourly cleaning cycle.
Applications
- Mixed Liquor
- Returned Activated Sludge
- ASP Control
Features
- Automatic Self Cleaning
- Low Cost of Ownership
- Large Sample Volume
Technical Specification
Physical
| Dimensions | 80 mm diameter, 540 mm long |
| Weight | 2.2 kg |
| Protection Class | IP68 |
| Material of Construction | Black Acetal Co-Polymer, 316 Stainless Steel, Glass |
| Seal Material | Polyurethane, Nitrile (Optional Viton) |
| Cable Entries | Integral Cable Gland |
| Cable Type | 4 core, 2 Twisted Pair, 5mm O/D Polyurethane Coated |
| Cable Length | 10 metres standard, 100 metres maximum |
| Service Requirement | Seal Service every 3500 cleans (application dependent) |
Environmental Data
| Operating Temperature | 0 - 50ºC |
| Storage Temperature | -20 to 60ºC |
| Location | Indoor or Outdoor |
| Pressure Rating | 10 mWC |
Interface to Monitor
| Sensor Output | PWM |
| Power | 12 VDC from Monitor |
Measurement Principal
| Principal of Operation | Light Absorption |
| Wavelength | 860 nm Infrared |
Measurement Characteristics
| Accuracy | Better than +/- 5% Reading on real sample |
| Resolution | Dependent on range setting, typcially +/- 2% |
| Repeatability | Better than +/- 1% Reading on real sample |
| Response Time | 0.5 seconds - damping provided by monitor |
| Flowrate | Not affected by flowrate, avoid deadspots and extreme turbulence |
| Maximum Range | 0-20,000 mg/l |
| Minimum Range | 0-4,000 mg/l |
| Measuring range will depend on the nature of the sample being monitored |

Download Turbi-Tech 2000 Manual
Download TurbiTech LA User List
Download Mounting Shaft Datasheet
Download Bracket Datasheet
Download Guide Tube Datasheet
Download Extension Bracket Datasheet

Share